HNRO7 quality comments - sample log sheets, console log sheets, chemical analysis log sheets July 15, 1999 (Lynne Talley) Codes: qflg = 4 bad value qflg = 3 suspicious value qflg = 2 good value Station 001 01 CTD oxygen very noisy CTD salinity spikes in strong T gradient Bottle 02 air leak before venting Autosal bad, switched after run Station 002 01 New CTD oxygen sensor installed before station CTD salinity spikes Autosal bad, switched after run 101 sal - too high, in gradient, suspicious (qflg = 3) ldt 106 sal - much too high, delete value (qflg=4) ldt Station 003 01 CTD salinity spikes 105 sal - too low (in constant S layer) (qflg = 4) ldt 106 sal - high, in gradient, probably OK (qflg = 2) ldt 107 sal - low, in gradient, probably OK (qflg=2) ldt Station 004 01 CTD - no comment Station 005 01 CTD salinity spikes 102 sal - sample drawn, not analyzed. 105 sal - too low, should be like 104, 103. No notes. (qflg=4) ldt 106 sal - high, in gradient, OK (qflg=2) ldt Station 006 02 Cast 01 - Mitchell bio-optics Double samples collected for Mitchell First sample fired well above bottom 212 sal - low, in gradient, OK (qflg=2) ldt 214 oxy - contaminated water, would not titrate. DLOG - duplicated 48 dbar level in 00602.bot for bottle #14. (console op pressed "b" too quick) Station 007 01 CTD salinity spikes on upcast 105 sal - low, in gradient, probably OK (qflg=2) ldt 106 sal - high, in gradient, probably OK (qflg=2) ldt Station 008 01 CTD - no comment. Bottle 11 air leak prior to venting, o-ring not seated properly. Station 009 01 CTD conductivity failed at about 270 m downcast. Cast aborted. DLOG - cracked conductivity sensor. Station 009 02 CTD 5 on, CTD 3 removed. O2 sensor from CTD 3 put on CTD 5. Bottle 108 salinity 0.2 low. Autosal log indicates rerun. Bad. (qflg=4) Bottle 109 salinity 0.2 low. Autosal log indicates rerun. Bad. (qflg=4) Station 010 01 CTD - no comment. Bottle 24 leaker (when stopcock pushed in and vent closed) Station 011 02 CTD - salinity noisy on upcast about 150m. Cast 1 - optics cast. Some double samples for optics. Station 012 02 CTD - no comment. Cast 1 - optics. Some double samples for optics. Station 013 01 CTD - no comment. Station 014 01 CTD - no comment. Station 015 01 CTD acquisition computer error at 25 m on upcast. Computer restarted as cast 2 and data set patched subsequently. DLOG - missed trip at 43.9dbar, got #'s from .scr Station 016 02 CTD - cast 1 optics. Some double samples for optics. SLOG - optic samplers found N23 had smaller water volume than N19 and N21 for this cast and a previous cast. Possibly tripped in the air? DLOG - 01602.bot only 23 trips. Duplicated surface trip values for 23 + 24 in h00 file. Station 017 02 CTD - salinity small but important difference order 0.002 between down and up which needs to be removed in calibration SLOG - Bottle 20 spigot pushed in on boarding. DLOG - only 23 trips. need to dup surface. Station 018 01 CTD - no comment. Station 019 01 CTD - no comment. Station 020 02 CTD - cast 1 optics. some double samples for optics. Station 021 01 CTD - no comment. Station 022 01 CTD - cast 2 was optics. some double samples on cast 1. DLOG - 8 bottle trips in bot, 14 bottles? Edited .bot file and duplicated missing pressure levels. Station 023 01 CTD - screen stopped updating before end of cast. Not clear if computer file is truncated. Station 024 01 CTD conductivity offset at 425-455 dbar downcast. Apparently returned to normal at 455. Station 025 01 CTD - no comment. DLOG - bottle 5 NO2 value same as bottle 6. (0.00) Station 026 01 Station 026 01 CTD - no comment. Station 027 01 CTD - VCR possibly not turned on. cast 2 was optics. some double samples on cast 1 SLOG - N23 closed in air. DLOG - Bottle 5 NO2 bad, delete. But save values in h00.sav Station 028 02 CTD - cast 1 was optics. some double samples on cast 2 DLOG - edited .bot file and duplicated missing press levels (~15,37,50). No duplicated from ... Station 029 01 CTD - no comment. SLOG - on N7 oxy was drawn after ph/alk. Station 030 01 CTD - no comment. SLOG - N24 leaking. Tripped in air? Station 031 01 CTD - no comment. Station 032 01 CTD - bottle 4 did not close. hung on conducting cable at pylon. SLOG - oxy flask 1442 was broken. used flask 1381 for 2nd draw. oxy on N18 - flask 1429 added 2ml MnCl2, added NaOH-NaI too late. probably bad. DLOG - bottle 3 looks like closed late. (nuts,oxy,freon look funny) bottle 4 did not close. bottle 5 NO2 same as bottle 6 value. (okay, but air bubble during peak) Station 033 02 CTD - cast 1 was optics. some double samples on cast 2. Station 034 02 CTD - cast 1 was optics. double surface bottle on cast 2. Station 035 01 CTD - no comment. Station 036 01 CTD - no comment. Station 037 01 CTD - no comment. SLOG - Nuts tube #12 empty. DLOG - nut on #12 not drawn. forgot. Station 038 01 CTD - cast 2 was optics. some double samples on cast 1. Station 039 01 CTD - drifter 15677 deployed after station. details in log. Station 040 02 CTD - cast 1 was optics. some double samples on cast 2. SLOG - air leak on N12. Station 041 01 CTD - drifter 15722 deployed after station. no details in log. Station 042 01 CTD - no comment. Station 043 01 CTD - probably problem with VCR tape. Station 044 01 CTD - cast 2 was optics. some double samples on cast 1. Station 045 02 CTD - cast 1 was optics. some double samples on cast 2. SLOG - spigot pushed in on N7. changed MnCl2 dispenser at N8. Station 046 01 CTD - no comment. SLOG - N15 changed interior spigot O-ring after sampling. Station 047 01 CTD - no comment. Station 048 01 CTD - no comment. Station 049 01 CTD - cast 2 was optics. some double samples on cast 1. Station 050 02 CTD - cast 1 was optics. some double samples on cast 2. SLOG - N5 leaking, vent not closed. Station 051 01 CTD - no comment. Station 052 01 CTD - no comment. Station 053 01 CTD - no comment. Station 054 01 CTD - no comment. Station 055 01 CTD - winch stop at 2400m on way up. Console op noticed bottom trip had not confirmed and 1st attempt at trip at 2400m failed confirm. Carl power cycled pylon box and got as confirm at 2400m. SLOG - N23 and N24 came up open. (not tripped) DLOG - tripping problem. Two bottom levels did not trip. Fixed in 05501.bot file. Station 056 01 CTD - no comment. Station 057 01 CTD - no confirm at initial attempt to trip bottom bottle. Power cycled pylon box and got confirm. Cast 2 was optics. some double samples on cast 1. DLOG - Extra bottom levels in 05701.bot removed. see 05701_sav for original file. Station 058 01 CTD - no confirm at bottom bottle, power cycle pylon box, 2nd no confirm on bottom bottle. Confirmation on 4th attempt. Cycled power 3 times. Station 059 01 CTD - no comment. DLOG - C2 and T2 now in place of C1 + T1 in data. (ie CTD sensors CTD5) Station 060 01 CTD - no comment. Station 061 02 CTD - cast 1 was optics. some double samples on cast 2. Station 062 02 CTD - cast 1 was optics. some double samples on cast 2. Station 063 01 CTD - no comment. Station 064 01 CTD - no comment. SLOG - N20 spigot pushed in. Station 065 01 CTD - on downcast stopped at 1427m. mucked up cond sensor. went up to 1478 [1378?] then continued down to bottom. upcast clean 1500-1200m. Station 066 01 CTD - no comment. DLOG - oxy in bottle 2 would not titrate. contamination in sample. Station 067 02 CTD - cast 1 was optics. some double samples on cast 2. SLOG - N7 through N11 in sunshine. Sampling jumped around. not simple 1 -> 24. Station 068 02 CTD - cast 1 was optics. some double samples on cast 2. SLOG - N12 leaking. open vent. Station 069 01 CTD - no comment. Station 070 01 CTD - cast 2 was optics. some double samples on cast 1. Station 071 02 CTD - cast 1 was optics. some double samples on cast 2. Probably touched bottom. bottom bottle tripped 6m above bottom. Station 072 01 CTD - no comment. Station 073 01 CTD - no comment. Station 074 01 CTD - no comment. Station 075 01 CTD - bottle 6 did not trip. bottle 6 did confirm. SLOG - N6 did not trip. Station 076 01 CTD - cast 2 was optics. some double samples on cast 1. bottle 6 did not trip. bottle 6 did confirm. Downcast stopped about 2650m to clear fouled cond sensor. Brought up about 50m then continued down. [LDT - bad data starts @ 2606, reversed up at 2682, back down at 2681] SLOG - N4 sampled by optics before tritium. N6 came up open. Station 077 02 CTD - cast 1 was optics. Station 078 01 CTD - no comment. Station 079 01 CTD - cast 2 was optics. some double samples on cast 1. Station 080 02 CTD - cast 1 was optics. some double samples on cast 2 Station 081 01 CTD - no comment. Station 082 01 CTD - no comment. Station 083 01 CTD - no comment. Station 084 01 CTD - no comment. Station 085 01 CTD - cast 2 was optics. some double samples on cast 1 Note 11/16/99: several salinities look bad on this station. ldt Station 086 01 CTD - no comment. Station 087 02 CTD - cast 1 was optics. some double samples on cast 2. Station 088 01 CTD - no comment. Station 089 01 CTD - no comment. Station 090 01 CTD - no comment. Station 091 01 CTD - no comment. Station 092 01 CTD - no comment. Station 093 01 CTD - cast 2 was optics. some double samples on cast 1. SLOG - N20 and N21 spigots pushed in on boarding. Station 094 02 CTD - cast 1 was optics. double sample at surface, cast 2. SLOG - N20 and N21 spigots pushed in on boarding. Replaced spigot inner O-rings after sampling. oxy sampling delayed on N2 (due to Heli tube prob). Station 095 01 CTD - no comment. SLOG - Not enough water in N3 and N13for tritium sample. Station 096 01 CTD - no comment. Station 097 01 CTD - no comment. Station 098 02 CTD - cast 1 was optics. some double samples on cast 2. bottle 6 did not trip. bottle 6 did confirm. SLOG - N6 came up open. Station 099 02 CTD - cast 1 was optics. some double samples on cast 2. altimeter failed. Station 100 01 CTD - no comment. SLOG - N16 spigot pushed in on boarding. Station 101 01 CTD - no comment. Station 102 01 CTD - no comment. Station 103 01 CTD - no comment. Station 104 01 CTD - no comment. SLOG - N6 vent open. oxy draw temps out of order N9 -> N13 Station 105 01 CTD - cast 2 was optics. some double samples on cast 1. Station 106 02 CTD - cast 1 was optics. some double samples on cast 2 Station 107 01 CTD - no comment. SLOG - N20 spigot pushed in on boarding. Station 108 01 CTD - no comment. Station 109 01 CTD - replaced oxy sensor prior to cast. new oxy sensor goes way off scale at ~2020m on downcast. Station 110 01 Station 111 01